AED Pro ZAS Software Download


AED Pro® ZOLL® Administrative Software (ZAS) can be used to update the software on your AED Pro to the latest revision or to adjust the default configuration.

AED Pro software version 2.84 or higher fully complies with the latest American Heart Association (AHA) and European Resuscitation (ERC) Guidelines.

How to determine the software version

AED Pro units manufactured after April 2011, when SW V2.84 was released, are fully compliant with the latest AHA and ERC Guidelines.

You can determine the manufacturing date of your AED Pro by referring to the two numbers and letter following the "AA" in the serial number. The two numbers following the AA indicate the year of manufacture, and the letter following the year indicates month of manufacture. For example, serial number AA12B123456 would have been manufactured in February of 2012.

What it takes

You need to have ZOLL Administrative Software (ZAS) v2.84 installed on your PC, and you will need an infrared port (IrDA device) to communicate with your AED Pro. With ZAS and an IrDA port you can easily upgrade the software in any AED Pro to v2.84 or adjust the device's configuration if needed.

To get started:

  • Download AED Pro ZAS v2.84 Upgrade Software after providing a valid serial number from an AED Pro.
  • Or contact your ZOLL Sales Representative or your Authorized ZOLL Distributor to order directly from ZOLL with a purchase order.

After upgrading your AED Pro:

You must register your upgrade with ZOLL to preserve your warranty and make certain you receive notification of any Device Corrective Actions that may occur in the future. Registration data are NOT forwarded to any other organization under any circumstances, and are NOT used for marketing purposes without your permission.